|
The customer reported to olympus that during a gastric bypass procedure, the jaw of this ultrasonic surgical device broke off and fell into the patient.The broken off device was retrieved from the patient, as reported.Attempts to obtain additional information were made, however, unsuccessful.There were no reports of patient or user harm associated with this event.
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The device was not returned to olympus.The reported phenomenon and condition of the device could not be confirmed.Therefore, it was not possible to identify which part of the device broke and fell off.Replication testing was carried out using an tb-0535fcs manufacturing by aomori (lot no.2xk 31).An excessive force was applied in direction for opening the grasping section.The grasping section detached from the insertion section.A force was further applied to grasping section in direction for opening.The distal end of the inner pipe deformed, but the grasping section did not fall off.The detached grasping section was rotated about two times in the twist direction by grasping the detached grasping section using equipment.The distal end of the inner pipe was twisted and torn, and the grasping section fell off.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the exact cause could not be determined because the circumstances surrounding the event (the reason why a force was applied to the grasping section) were unclear.However, based on the result of replication testing, a likely reason for the falling of the grasping section might be the following: 1.An excessive force in the direction for opening was applied to the grasping section.This caused the grasping section to detach from the shaft.2.Since a force in the twist direction was applied to the detached grasping section, the distal end of the inner pipe broke.As a result, the grasping section fell off.The event can be prevented by following the instructions for use which state: "when inserting the thunderbeat into or withdrawing it from the trocar tube, gently hold the control handle and make sure that the grasping section is closed.If the thunderbeat is inserted or withdrawn with the grasping section open, the probe tip/grasping section may become damaged, or it may become impossible to withdraw it from the trocar.Should any crack, scratch, deformation, split, protrusion, or partial separating be observed on the probe tip, grasping section, tissue pad, shaft, the surface of the transducer, transducer cord, or transducer plug, do not use them and replace the damaged instrument or the transducer with a spare.Using a damaged device may cause burns due to abnormal output or high-frequency (rf bipolar) current leakage or breakage of the probe tip, the tissue pad, and the grasping section." olympus will continue to monitor field performance for this device.
|