| Model Number |
HY11D10R4 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/02/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluation summary: visual inspection shows no outward signs of damage.There was a tie-wrap around the qb-in port.Performed pressure integrity testing, 0.5 l/pm with 23 psi, (1189 mmhg) of backpressure for 10 min.During the pressure integrity test there was a leak observed at the qb-in port.The leak is coming from the tubing/port connection.Conclusion: complaint is confirmed for component with leakage.After evaluation the cause of this complaint could not be determined since in the investigation nothing manufacturing-related can replicate the defect, the connection was made according the catalog and the raw material did not show any anomalies.Review of the device history record found no abnormalities during manufacturing that would cause or contribute to the reported event.There were no adverse patient effects as a result of this incidence.Medtronic will continue to monitor for future occurrences and trends.This regulatory report is being submitted as part of a retrospective review and remediation per d00953163 as part of a capa action.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information that prior to use of a custom tubing pack, the customer reported that after setting up the extracorporeal circuit, priming, and beginning recirculation, the myotherm cardioplegia heat exchanger leaked profusely during pressure testing at 250 mmhg.The device was replaced to complete the procedure.There was no patient involvement so no adverse effect occurred.Additional information: the pack tray was undamaged, and the outside box was not available.
|
| |
|
Search Alerts/Recalls
|
|
|