• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 RECLAIM ASSEMBLY BODY; RECLAIM INSTRUMENTS : INSERTION DEVICES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 RECLAIM ASSEMBLY BODY; RECLAIM INSTRUMENTS : INSERTION DEVICES Back to Search Results
Model Number 2975-00-600
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/30/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the device fits super tight with another piece that is intended to make the torque with the sand watch.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: the device associated with this report was not returned to depuy synthes for evaluation, however photos were provided for review.Review of the provided photo just shows that the pull rod is inserted inside the reclaim assembly body, but the reported allegation of tightness or jammed condition of the devices could not be confirmed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : the piece ref d297500600 was super tight together with another piece that are used to make the torque with the sand watch.The piece is marked with red tape and there are photos in the attached document.The device associated with this report was returned to depuy synthes for evaluation.Visual inspection of the returned sample was able to confirm the reported jammed condition.A functional test to assess the allegation of jammed was conducted but it was not possible to twist the pull rod cap was not.The complaint condition was replicated.No other problem identified.The potential cause can be attributable to excessive force at the moment of rotating the pull rod cap.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RECLAIM ASSEMBLY BODY
Common Device Name
RECLAIM INSTRUMENTS : INSERTION DEVICES
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY IRELAND 9616671
loughbeg, ringaskiddy co.
cork
EI  
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key17097889
Report Number1818910-2023-11960
Device Sequence Number7690308
Product Code LXH
UDI-Device Identifier10603295155942
UDI-Public10603295155942
Combination Product (Y/N)N
Initial Reporter CountryCO
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 06/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2975-00-600
Device Catalogue Number297500600
Device Lot NumberSO2039473
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/01/2023
Supplement Date Received by Manufacturer06/16/2023
08/03/2023
08/29/2023
Initial Report FDA Received Date06/09/2023
Supplement Report FDA Received Date06/19/2023
08/04/2023
08/29/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/11/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-