Model Number N/A
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Failure of Implant (1924); Pain (1994); Swelling/ Edema (4577)
Date of Event 05/22/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported patient underwent a revision procedure five months post implantation due to tibial loosening.Patient had ongoing pain and swelling of the knee after initial six week post-op period.Attempts to obtain additional information have been made; however, no more is available at this time.
 
Additional Manufacturer Narrative
(b)(4).D10 - medical product: femur cemented cruciate retaining (cr) catalog # 42502607401 lot # 65396593 articular surface medial congruent (mc) use with tibia catalog # 42512101010 lot # 64257601.G2: australia.H3: customer has indicated that the product will not be returned because product location is unknown.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.H3 other text : product location unknown.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4, b5, d9, g3, g6, h1, h2, h3, h4, h6, h10 reported event was confirmed by review of x-rays provided.Visual evaluation of the returned device shows signs of implantation with scratches, nicked, and foreign material on the distal surface.Device history record (dhr) was reviewed and no discrepancies relevant to the reported event were found.A definitive root cause cannot be determined.X-ray review indicates there is femoral notching is present.Minimal lucency also seen along the cement hardware interface of the femoral component on the lateral film.Moderate joint effusion.Ossific density within the posterior joint space could represent fabella versus loose body.Overall fit and alignment of the implants is appropriate.Radiolucency along the hardware indicating possible loosening of the femoral and tibial components.Suggestion of minimal cement on the ap film involving the medial aspect of the tibial component.More prominent appearance of cement within the central aspect of the distal femur.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NATURAL TIBIA TRABECULAR METAL TWO-PEG POROUS FIXED BEARING
Common Device Name
PROSTHESIS KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key17097949
Report Number0001822565-2023-01540
Device Sequence Number7826322
Product Code MBH
UDI-Device Identifier00889024510616
UDI-Public(01)00889024510616(17)291031(10)64458985
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K173417
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number42530008301
Device Lot Number64458985
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/19/2023
Supplement Date Received by Manufacturer10/26/2023
Initial Report FDA Received Date06/09/2023
Supplement Report FDA Received Date10/26/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/24/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10 NARRATIVE.
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexMale
Patient Weight105 KG
-
-