Product event summary: the afapro28 balloon catheter with lot number 18484 was returned and analyzed.External visual inspection of the balloon segment showed a leak path from the outer balloon.The outer balloon proximal bond was partially detached from the catheter shaft.External visual inspection of the balloon segment showed blood/fluid inside the balloon.During inspection of the handle segment, blood/ liquid was observed inside handle.The catheter smart chip data was downloaded and reviewed.Data indicated the catheter was used for 17 applications on the reported event date.During functional testing, the terminated the application and triggered system notice 50006 "the safety system detected blood in the catheter handle, the injection was stopped and the vacuum disabled." pressure testing and inspection of sub-components was performed on the balloon, handle, and shaft segments.Pressurizing of the outer balloon identified the outer balloon proximal bond was leaking and detached from the shaft.During inspection of the shaft segment, a guide wire lumen kink/twist was observed 1.5 inches proximal to the catheter tip.Pressure testing did not identify any leakage from the kink.In conclusion, the balloon catheter failed the returned product inspection due to the guide wire lumen kink and deta ched guide wire lumen.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|