• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE REVISION TIBIAL BLOCK L; RESECTION BLOCKS/DEVICES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE REVISION TIBIAL BLOCK L; RESECTION BLOCKS/DEVICES Back to Search Results
Model Number 2506-10-111
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/30/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the femoral trial that has a crack at the notch.There was no surgical delay.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).H6:part/ component/ sub assembly term not applicable (g07001).Investigation summary : the device associated with this report was returned to depuy synthes for evaluation.Visual analysis of the device found a small burr at the 10° drill hole.The allegation can be confirmed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE REVISION TIBIAL BLOCK L
Common Device Name
RESECTION BLOCKS/DEVICES
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY IRELAND 9616671
loughbeg, ringaskiddy co.
cork
EI  
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key17099434
Report Number1818910-2023-11998
Device Sequence Number15410378
Product Code HTZ
UDI-Device Identifier10603295436225
UDI-Public10603295436225
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 06/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2506-10-111
Device Catalogue Number250610111
Device Lot NumberSO2032245
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/29/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/06/2023
Initial Report FDA Received Date06/09/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/22/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-