| Model Number |
Z508G |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/23/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a patient underwent unknown plastic surgery on (b)(6) 2023 and suture was used.Some needles were broken at the middle part during interrupted suture on dermis.After that, some needles were broken at the needle tip part.This happened in 5 needles in total.There were no adverse consequences to the patient.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.Did the needle fall into the patient? if yes.Was the needle piece(s) retrieved during the same procedure? what tissue structure the broken needle was located? were x-rays taken to locate the needle piece(s)? what measures were taken to retrieve the broken piece? was there any additional tissue damage as a result of searching for the needle piece? attempts are being made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.H6 component code: g07002 ¿ device not returned.Related events captured via 2210968-2023-04217, 2210968-2023-04218, 2210968-2023-04219 and 2210968-2023-04220.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Date sent to the fda: 6/26/2023.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Additional information was requested, the following was obtained: did the needle fall into the patient? no.Please perform and document the follow up attempt for product return.Please document the shipment tracking number in notes or rmao sections.The device has been shipped.Please check rmao.Please provide the source or name of person providing answers to follow-up questions (not the person relaying/submitting answers to loc or chu).(b)(6).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Related reports: 2210968-2023-04216, 2210968-2023-04217, 2210968-2023-04218, 2210968-2023-04219 and 2210968-2023-04220.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Date sent to the fda: 7/10/2023.H6 component code: g07002 no device problem found.H3 investigational summary: the product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.The returned sample revealed that it was received, one opened sample that pertain to product code z508g.In order to evaluate the condition of the returned sample, the packet was opened.The swage and attachment area were noted to be as expected.The tip and body of the needle were examined under magnification and no defects or needle breakage were found.In addition, the sample was tested by resistance test to the needle and met the requirements.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.The event described could not be confirmed as the device performed without any difficulties noted.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.Additional h3 investigational summary: the product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.The returned sample revealed that it was received, one open box with seven unopened samples that pertain to product code z508g.In order to evaluate the condition of the returned samples, the packets were opened.The swage and attachment area were noted to be as expected.The tip and body of the needles were examined under magnification and no defects or needle breakage were found.In addition, the samples were tested by resistance test to needle and met the requirements.As part of ethicon quality process all devices are manufactured, inspected, and released to approved specifications.The event described could not be confirmed as the device performed without any difficulties noted.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Related reports: 2210968-2023-04217, 2210968-2023-04218, 2210968-2023-04219 and 2210968-2023-04220.
|
| |
|
Search Alerts/Recalls
|
|