| Model Number |
LOW PROFILE SCREW, 1.4X11MM, CORTICAL,TI |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
05/18/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2023, it was reported by a sales representative via email that an ar-18714-11 low profile screw, an ar-18714-12 low profile screw, and an ar-18714v-11 va locking screw broke during insertion into the plate.This was discovered during a metacarpal fx procedure on (b)(6) 2023.The screws remain inside the patient.Additional information requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to user-applied excessive force during the tightening process.
|
| |
|