C.R. BARD, INC. (BASD) -3006260740 SAFESTEP HUBER NEEDLE SET 22G X 0.5IN WITH Y-SITE; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number LH-0035YN
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/28/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The manufacturer has received the sample and will evaluate.Results are expected soon.
 
Event or Problem Description
It was reported by the customer when the device was primed with saline solution prior to use, air was aspirated and air contamination was found.There was no reported patient involvement.It was reported this occurred with two devices.This report addresses the second device.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.Based on a review of this information, the following was concluded: the complaint of a leak path in the returned infusion sets is unconfirmed because the problem could not be reproduced.Two 22 ga x 0.5 in.Safestep infusion sets with y-sites were returned for evaluation.During functional testing of each luer for each returned infusion set, pressurization of the devices during infusion of water using a 12 ml syringe revealed no leaks throughout the returned safestep infusion sets.Water was aspirated into each device with no obvious leaking of air.Because leaks in the infusion sets could not be found during functional testing, the complaint of a leaking infusion sets is unconfirmed.This complaint will be recorded for future trending and monitoring purposes.H3 other text : evaluation findings are in section h.11.
 
Event or Problem Description
It was reported by the customer when the device was primed with saline solution prior to use, air was aspirated and air contamination was found.There was no reported patient involvement.It was reported this occurred with two devices.This report addresses the second device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SAFESTEP HUBER NEEDLE SET 22G X 0.5IN WITH Y-SITE
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer Contact
suedabba mahboobi
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key17102123
Report Number3006260740-2023-02389
Device Sequence Number16627428
Product Code FPA
UDI-Device Identifier00801741066290
UDI-Public(01)00801741066290
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K153440
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberLH-0035YN
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/07/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/23/2023
Supplement Date Received by Manufacturer10/11/2023
Initial Report FDA Received Date06/10/2023
Supplement Report FDA Received Date10/16/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-