| Model Number |
420184-14 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/29/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) did receive a da vinci product to perform failure analysis.The pcs instrument was analyzed and found to have a broken pitch cable at the distal end.The broken cable segment that contains the crimp was still installed in the clevis.Common causes of this failure mode are attributed to a component failure.When pitch cable breakage occurs, the instrument is immediately inoperable due to loss of functionality.The instrument has tungsten drive cables to transmit motion from the input discs in the housing, through the main shaft, and to the distal instrument tip.These cables are exposed at the instrument tip and control movements at the wrist.
|
| |
|
Event or Problem Description
|
|
It was reported that during a da vinci-assisted myomectomy surgical procedure, there was a stopping point when the angle of the permanent cautery spatula instrument changes from 60 degrees on the left to 60 degrees on the right.The instrument was also identified with a broken cable.The procedure was completed with no reported injury.
|
| |
|