STRYKER NEUROVASCULAR CORK NEUROFORM ATLAS 3.0MM X 21MM NO TIP- PMA; STENT, INTRACRANIAL NEUROVASCULAR Back to Search Results
Model Number M003UZAS30210
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Codes Obstruction/Occlusion (2422); Foreign Body In Patient (2687)
Date of Event 05/22/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H3 other text : the device is not available to the manufacturer.
 
Event or Problem Description
It was reported that during the procedure, after deploying the subject stent in the m2/m1 vessel the operator tried to re-cross the wire to select the other vessel for the second stent.The guidewire got stuck on to the subject stent.While trying to free the wire, the operator pulled out a portion of the subject stent.The patient¿s m1 vessel appeared to be occluded in the final run.No further information available.
 
Event or Problem Description
It was reported that during the procedure, after deploying the subject stent in the m2/m1 vessel the operator tried to re-cross the wire to select the other vessel for the second stent.The guidewire got stuck on to the subject stent.While trying to free the wire, the operator pulled out a portion of the subject stent.The patient¿s m1 vessel appeared to be occluded in the final run.Additional information received on 23-june-2023 confirmed that a wide neck aneurysm in the middle cerebral artery (mca) bifurcation was being treated but the procedure was not completed.The physician attempted to remove the broken section of the stent with a snare but was unsuccessful.No further information available.
 
Additional Manufacturer Narrative
Due to the automated manufacturing execution system (mes) system there are controls in the manufacturing process to ensure the product met specifications upon release.Visual and functional inspections were not performed due to the product was not returned.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported event was unable to be confirmed and it cannot be confirmed that the device met specification, as the device was not returned.As the product was not returned and review and analysis of all available information fails to indicate an assignable cause or probable assignable cause for the reported event, an assignable cause of undeterminable will be assigned to the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEUROFORM ATLAS 3.0MM X 21MM NO TIP- PMA
Common Device Name
STENT, INTRACRANIAL NEUROVASCULAR
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI  NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI   NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key17104766
Report Number3008881809-2023-00294
Device Sequence Number6069979
Product Code NJE
UDI-Device Identifier07613327382211
UDI-Public07613327382211
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P180031
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberM003UZAS30210
Device Catalogue NumberM003UZAS30210
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/22/2023
Supplement Date Received by Manufacturer06/29/2023
Initial Report FDA Received Date06/12/2023
Supplement Report FDA Received Date07/26/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
EXCELSIOR SL-10 MICROCATHETER (STRYKER); SYNCHRO SELECT GUIDEWIRE (STRYKER)
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient SexUnknown
-
-