| Catalog Number |
82005 |
| Medical Device Problem Code |
Overheating of Device (1437)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/05/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Since an overheating insert could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
|
| |
|
Event or Problem Description
|
|
In this event it is reported that a 30k fsi-sli-fg-10s ins,pkd overheated during use.No injury occurred.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: adding udi # (b)(4).This is a follow up report for this information.Returned qty.1 insert, 82005, 23012-00089725, jz, warranty test method / gp005-wi02 inductance: gauge id# 5349 due: 11/30/2024 result: 3.22 meets spec does not meet spec tip condition: - meets spec does not meet spec stack condition meets spec does not meet spec tested on: g136 gage id# 9626-2 due date: 03/31/2023 set pressure: 35 psi water flow: output rate: 35 psi - meets spec does not meet spec spray pattern: - meets spec does not meet spec water leak: hp sealing ring proximal ring distal ring seam leak vibration: - meets spec does not meet spec grip condition: meets spec does not meet spec other visual observation: insert tip is broken at edm hole.Alleged event: confirmed - not confirmed.
|
| |
|
Search Alerts/Recalls
|