• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INTERLOCK-35; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INTERLOCK-35; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION Back to Search Results
Model Number 83786
Medical Device Problem Code Premature Activation (1484)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/31/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device evaluated by manufacturer: the device was returned for analysis.The main coil return inside a non-boston scientific catheter, a torque device also returns.The main coil was observed bent, detached and stretched at the coil arm section.No more damages were observed.Microscope inspection revealed that under the microscope it was observed that the main coil was observed bent, detached and stretched at the coil arm section.Functional inspection.The functional test could not be performed due only the main coil returned.Dimensional inspection revealed that the overall dimension of the zap tip, primary coil and the coil arm of the main coil were within the specification.
 
Event or Problem Description
Reportable based on device investigations completed on (b)(6) 2023.It was reported that premature deployment occurred.The target lesion was located in the varicose veins of the fundus of the stomach with severe tortuosity.A 6mm x 20cm interlock.035 was selected for use.During the procedure, it was noted that the interlocking arm of the coil detached/unlocked from the interlocking arm of the pusher wire.The coil and the catheter were removed directly as a whole and the procedure was completed with another of the same device.There were no complications and the patient is stable.However, device analysis revealed that the coil was detached at the coil arm section.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERLOCK-35
Common Device Name
DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORK LIMITED
model farm road
cork
EI  
Manufacturer Contact
jeff wallner
4100 hamline ave n
arden hills, MN 55112
6515811560
MDR Report Key17107066
Report Number2124215-2023-25262
Device Sequence Number11464373
Product Code KRD
UDI-Device Identifier08714729793021
UDI-Public08714729793021
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K133208
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/01/2025
Device Model Number83786
Device Catalogue Number83786
Device Lot Number0030074228
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/08/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/15/2023
Initial Report FDA Received Date06/12/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/02/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MICROCATHETER-CORDIS 5F 0.038 INCH BORE
Patient Age71 YR
Patient SexFemale
Patient Weight63 KG
-
-