MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE RP TIB BASE SZ 6 POR; ATTUNE CEMENTLESS IMPLANTS : KNEE TIBIAL TRAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE RP TIB BASE SZ 6 POR; ATTUNE CEMENTLESS IMPLANTS : KNEE TIBIAL TRAY Back to Search Results
Model Number 1506-11-006
Medical Device Problem Code Device Contaminated During Manufacture or Shipping (2969)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthese has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthese or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the inner sterile package had a hair under seal on foam insert.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary the device was not returned to depuy synthes for evaluation, however photos were provided for review.Review of the provided photos revealed a hair under the seal over the foam.The report allegation can be confirmed, however is not possible to determine the open condition of the sterile packaging.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot the device lot number is unknown, therefore a device history review could not be performed.If more information become available, the record will be re-assessed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE RP TIB BASE SZ 6 POR
Common Device Name
ATTUNE CEMENTLESS IMPLANTS : KNEE TIBIAL TRAY
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key17107204
Report Number1818910-2023-12075
Device Sequence Number16852409
Product Code NJL
UDI-Device Identifier10603295416289
UDI-Public10603295416289
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
P830055
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1506-11-006
Device Catalogue Number150611006
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/05/2023
Supplement Date Received by Manufacturer06/28/2023
Initial Report FDA Received Date06/12/2023
Supplement Report FDA Received Date06/29/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-