| Model Number |
00884450020135 |
| Medical Device Problem Codes |
Fracture (1260); Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/05/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The suspect device is expected to return for evaluation.A follow up will be submitted when the evaluation is complete.
|
| |
|
Event or Problem Description
|
|
The account alleges that an aorto-gram of the left common iliac/external iliac artery was performed when the catheter tip detached.The foreign body was retrieved successfully with a vascular snare device.
|
| |
|
Additional Manufacturer Narrative
|
|
The suspect medical device has been returned for a full evaluation.The device was visually and microscopically investigated.The complaint is confirmed.The root cause is related to the clinical forces imposed upon the device within a tortuous anatomy.The device history record was reviewed, and no exception documents were found.A search of the complaint database was performed and no similar complaints for this lot number were identified.
|
| |
|