STRYKER ORTHOPAEDICS-MAHWAH NO 3. TRIATHLON TS PLUS TIBIAL INSERT X3 POLY 9MM; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
|
Back to Search Results |
|
| Model Number |
5537-G-309-E |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Impaired Healing (2378)
|
| Date of Event |
05/24/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient's left knee was revised.As reported: "patient had a poly swap due to incision not fully healed.
|
| |
|
Additional Manufacturer Narrative
|
|
An event regarding patient factors (impaired healing) involving a triathlon insert was reported.The event was not confirmed.Method & results product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.Clinician review: no medical records were received for review with a clinical consultant.Product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient had a poly swap due to incision not fully healed the exact cause of the event could not be determined because insufficient information was provided.Further information such as return of the device, pathology reports, pre- and post-operative x-rays and the primary operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and / or additional information become available to indicate further evaluation is warranted, this record will be reopened.H3 other text : not returned to the manufacturer.
|
| |
|
Search Alerts/Recalls
|
|
|