| Model Number |
03.58.443 |
| Medical Device Problem Code |
Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/16/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Batch review performed on 19 may 2023
lot 2222516: 30 items manufactured and released on 13-jun-2022.Expiration date: 2027-may-24.No anomalies found related to the problem.To date, 26 items of the same lot have been sold with no similar reported event during the period of review.Mysolution planning analysis:
our analysis of the myspine process of this case found no deviations from the standard procedures.Each step has been performed correctly.
|
| |
|
Event or Problem Description
|
|
The patient came in for a post-op appointment and it was observed from x-rays that a screw had become misplaced.At 5 days after the primary surgery, the surgeon redirected the screw, and the surgery was completed successfully.Fluoroscopy was used during the primary.The screw was likely malpositioned during the primary surgery.No issue with the myspine guides occurred during the primary surgery.
|
| |
|