| Model Number |
1100-10-100 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Bone Fracture(s) (1870); Pain (1994)
|
| Date of Event |
05/27/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that this was a tsa performed on (b)(6) 2023.The surgery was completed successfully without any surgical delay.On (b)(6) 2023, the patient complained of pain.On (b)(6) 2023, it was confirmed that the greater tubercle was displaced around the top of the head.Therefore, the surgeon decided to perform revision surgery instead of preservation.On (b)(6) 2023, a revision surgery was performed.The surgeon assumed the worst and started the surgery with the possibility of replacing rsa.Only the body and head were replaced.It was locked with soft anchor.A burr hole was created in shaft.Permatepe and dynatape were placed over the rotator cuff and inserted into the burr hole, holding the greater tubercle, in placed and locked.According to the surgeon, other possible causes besides the products are the angle of the posterior twist and the greater tubercle injury during the retractor operation.No further information is available.Doi: (b)(6) 2023; doe: (b)(6) 2023; affected side: unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).H6 component code: appropriate term/code not available (g07002) used to capture no findings available.Investigation summary - no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|