• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER NEUROVASCULAR CORK TARGET HELICAL ULTRA 4MM X 8CM; DEVICE, NEUROVASCULAR EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER NEUROVASCULAR CORK TARGET HELICAL ULTRA 4MM X 8CM; DEVICE, NEUROVASCULAR EMBOLIZATION Back to Search Results
Model Number M0035434080
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/17/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H3: the device is not available to the manufacturer.
 
Event or Problem Description
It was reported that at the time of release the coil (subject device), it was partially released, remaining adhered to the delivery system, which was left outside the aneurysmal sac, rupturing between the struts of the stent and the vessel wall.The procedure was completed successfully.No clinical consequences were reported to the patient due to this event.
 
Event or Problem Description
It was reported that at the time of release the coil (subject device), it was partially released, remaining adhered to the delivery system, which was left outside the aneurysmal sac, rupturing between the struts of the stent and the vessel wall.The procedure was completed successfully.No clinical consequences were reported to the patient due to this event.
 
Additional Manufacturer Narrative
H4 manufacturing date ¿ added d4 expiration date - added due to the automated mes (manufacturing execution system) system there are controls in the manufacturing process to ensure the product met specifications upon release.The subject device was not available; therefore, a visual inspection as well as a functional evaluation could not be performed.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported complaint could not be confirmed, and it could not be definitively determined if the device failed to meet specifications because the product was not returned.While there are a number of potential causes for the reported issue, because review and analysis of available information failed to identify a definitive cause and the device was not returned, an assignable cause of undeterminable was assigned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TARGET HELICAL ULTRA 4MM X 8CM
Common Device Name
DEVICE, NEUROVASCULAR EMBOLIZATION
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI  NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI   NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key17119173
Report Number3008881809-2023-00298
Device Sequence Number15213532
Product Code HCG
UDI-Device Identifier04546540676115
UDI-Public04546540676115
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K153658
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/21/2025
Device Model NumberM0035434080
Device Catalogue NumberM0035434080
Device Lot Number24030862
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/01/2023
Supplement Date Received by Manufacturer08/15/2023
Initial Report FDA Received Date06/13/2023
Supplement Report FDA Received Date08/24/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/23/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
STENT (UNKNOWN)
Patient Age43 YR
Patient SexFemale
-
-