MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 MINILOC SAFETY INFUSION SET 19G X 1 IN WITH Y-SITE; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 MINILOC SAFETY INFUSION SET 19G X 1 IN WITH Y-SITE; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number N/A
Medical Device Problem Code Component Missing (2306)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/11/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported by the customer port needle exposed in packaging.Needle missing protective cover when package opened.No one punctured.No other information was provided.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The device has not been returned to the manufacturer for evaluation.H3 other text : device not returned for evaluation.
 
Event or Problem Description
It was reported via medwatch "port needle exposed in packaging.Needle missing protective cover when package opened.No one punctured." no other information was provided.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The complaint of a missing needle protective cover is inconclusive due to poor sample condition.One 19 ga x 1 in miniloc infusion set was returned in open packaging indicating lot: asgzfc145.The safety mechanism was not activated.The needle protective cover was not observed on the device or within the open packaging.Microscopic inspection of the clear packaging film revealed what appeared to be a puncture.The condition of the device and packaging prior to handling and opening is unknown.Based on the information provided, possible contributing factors include dislodgement of the needle cover during handling and omission during assembly.Since the condition of the device prior to opening remains unknown, the complaint could not be confirmed and remains inconclusive at this time.H3 other text : evaluation findings are in section h.11.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MINILOC SAFETY INFUSION SET 19G X 1 IN WITH Y-SITE
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer Contact
johanna de oliveira
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key17121386
Report Number3006260740-2023-02411
Device Sequence Number17790508
Product Code FPA
UDI-Device Identifier00801741047985
UDI-Public(01)00801741047985
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K050600
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Risk Manager
Type of Report Initial,Followup
Report Date (Section B) 09/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/30/2025
Device Model NumberN/A
Device Catalogue Number0681910
Device Lot NumberASGZFC145
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/25/2023
Supplement Date Received by Manufacturer09/06/2023
Initial Report FDA Received Date06/13/2023
Supplement Report FDA Received Date09/08/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/01/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-