| Model Number |
SRX-700 |
| Medical Device Problem Codes |
Device Appears to Trigger Rejection (1524); Material Rupture (1546)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
05/30/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture baker grade iii.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side capsular contracture baker grade iii.Device remains implanted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional later reported right-side rupture discovered during surgery.Device has been explanted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional later reported right-side rupture discovered during surgery.Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the device analysis grid, the assessments of the complaint are: ¿ rupture : observed, opening on the anterior side assessed as surgical damage.¿ capsular contracture : unable to observe since it is a medical event and is not related to the device.No additional observations.¿ crease fold observed on the device.No further actions are required as no issues with the manufacturing process are observed.
|
| |
|
Search Alerts/Recalls
|