MAUDE Adverse Event Report: BIOMET 3I OSSEOTITE IMPLANT 4 X 11.5MM; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET 3I OSSEOTITE IMPLANT 4 X 11.5MM; DENTAL IMPLANT Back to Search Results
Catalog Number OSS411
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Codes Pain (1994); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/22/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).A4: weight unknown / not proided.
 
Event or Problem Description
It was reported implant fracture in the cervical third at tooth site 26.Abutment screw is intact.Patient referred pain and was rescheduled for removal of the remaining part of the implant.
 
Additional Manufacturer Narrative
Zimmer biomet complaint number (b)(4).Zimvie received one (1) oss411, (osseotite® implant 4 x 11.5mm) for evaluation.Visual evaluation of the as returned device identified the implant with signs of use, and apparent bone / tissue attached to external threads.The implant was fractured at the collar region.Functional testing to recreate the reported event could not be performed due to the nature of the device and event.This complaint refers to the specific device being investigated for this complaint record.Device history record (dhr) review was completed for the subject lot number 2016010214.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number 2016010214 for similar events and no other complaint was identified.Review completed utilizing keywords: fracture implant.Based on the investigation and risk management file review, the most likely root causes determined from the investigation are missing or confusing instructions for use, clinician selects implant that is unable to resist long-term occlusal loading, clinician places implant in a patient with patient factors (e.G.Bruxism) incompatible with long-term osseointegration of implant.Therefore, based on the available information, a device malfunction did occur.The reported event was confirmed with all the available information.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the product was nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.At this time, the complaint investigation has been completed and the record will be closed.If additional information is received, the record will be re-opened for further evaluation.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OSSEOTITE IMPLANT 4 X 11.5MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
susanne taylor
4555 riverside drive
palm beach gardens, FL 33410
5617766700
MDR Report Key17123327
Report Number0001038806-2023-01144
Device Sequence Number6458250
Product Code DZE
UDI-Device Identifier00844868013181
UDI-Public(01)00844868013181(17)210120(10)2016010214
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K063286
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 12/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date01/19/2021
Device Catalogue NumberOSS411
Device Lot Number2016010214
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/24/2023
Supplement Date Received by Manufacturer05/24/2023
Initial Report FDA Received Date06/14/2023
Supplement Report FDA Received Date12/11/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/19/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age44 YR
Patient SexPrefer Not To Disclose
Patient EthnicityNon Hispanic
-
-