Model Number 627644
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Loss of Range of Motion (2032); Nonunion/Delayed-union Bone Fracture (2369); Implant Pain (4561)
Date of Event 05/19/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Based on the available information the device will not be returned therefore an evaluation of the device cannot be performed.Should additional information become available, it will be provided in a supplemental report.H3 other text : hospital retained device.
 
Event or Problem Description
It was reported that the patient's right femur was revised.After a previous orif to address a femoral fracture, patient presented complaining of pain and limited rom.X-ray showed non-union of the bone with 3 broken screws and one screw which pulled out with the plate.A femoral stem and head, the plate and screws, and a competitor knee were revised to a competitor total femur replacement (acetabular components were left in situ).Rep confirmed that no further information will be released by the hospital or surgeon.
 
Additional Manufacturer Narrative
Please note the correction to b1.The reported event could not be confirmed, since the device was not returned for evaluation and no other evidences were provided.The device inspection was not possible as the product was not returned for investigation.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.An x-ray was provided which were presented to our health care professional for evaluation, and a question about the most likely reason for this event, to which the medical expert replied: ¿it is likely to be related to a failure in stability from the entire construct leading to the non-union.I can say some things about the construction.Proximal screws are missing in the construction.However, since i don¿t have the trauma x-rays, the post-operative pics, etc., so there is no complete picture.¿ more detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.If device is returned or any further information is provided, the investigation report will be reassessed.
 
Event or Problem Description
It was reported that the patient's right femur was revised.After a previous orif to address a femoral fracture, patient presented complaining of pain and limited rom.X-ray showed non-union of the bone with 3 broken screws and one screw which pulled out with the plate.A femoral stem and head, the plate and screws, and a competitor knee were revised to a competitor total femur replacement (acetabular components were left in situ).Rep confirmed that no further information will be released by the hospital or surgeon.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISTAL LATERAL FEMUR PLATE AXSOS 3 FOR RIGHT FEMUR 14 HOLE / L310MM
Common Device Name
PLATE, FIXATION, BONE
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
Manufacturer (Section G)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ   2545
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key17124169
Report Number0008031020-2023-00229
Device Sequence Number17773097
Product Code HRS
UDI-Device Identifier07613252579687
UDI-Public07613252579687
Combination Product (Y/N)N
Initial Reporter StateRI
Initial Reporter CountryUS
PMA/510(K) Number
K222381
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2023
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number627644
Device Catalogue Number627644
Device Lot NumberH24832
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/19/2023
Supplement Date Received by Manufacturer06/28/2023
Initial Report FDA Received Date06/14/2023
Supplement Report FDA Received Date07/25/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/24/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age83 YR
Patient SexFemale
Patient Weight84 KG
-
-