| Model Number |
0014A |
| Medical Device Problem Code |
Manufacturing, Packaging or Shipping Problem (2975)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that not during a procedure that all the packing and labeling was correct but the incorrect device included.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 6/14/2023.B3: only event year known: 2023.Investigation summary the product was returned for evaluation.Visual inspection was conducted on the returned device.Forty electrodes were returned and 13 were sampled for analysis.Visual analysis of the returned sample revealed that the 0014a device (b) was received inside its sterile package.Upon visual inspection, it was noted that the label is for a 0014a e-z clean blade 4.0 inch; however, inside it has a 6.5-inch e-z clean blade.As part of our quality process, the manufacturing records of this lot number were reviewed, and the manufacturing standards were met prior to the release of this lot.This defect has been correlated to a manufacturing process.Based on the information currently available, a product issue was identified during the investigation of the sample received.This product issue will be addressed through ethicon endo surgery¿s quality system.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 6/15/2023 photo analysis: this is an analysis of an image submitted for evaluation.Image: the image provided by the customer shows a 0014a e-z clean blade 4.0 inch inside of 6.5-inch sterile package.Based on the information currently available, a product issue was identified during the investigation of the sample received. this product issue will be addressed through the ethicon endo-surgery quality system.
|
| |
|
Search Alerts/Recalls
|