| Catalog Number |
03P89-24 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
Valvular Stenosis (2697)
|
| Date of Event |
05/25/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On (b)(6) 2023, abbott point of care was contacted by a customer regarding i-stat pt/inr cartridges that yielded suspected discrepant inr results a 53 year old female patient seen for a mechanical valve replacement.There was no additional patient information available at the time of this report.Return product is available for investigation.Method date collected tested inr sample i-stat (b)(6) 2023 02:23 02:23 4.0 whole blood- a.Stago (b)(6) 2023 02:02 ni 2.6 plasma- b.Pt/inr, intended use the i-stat pt, a prothrombin time test, is useful for monitoring patients receiving oral anticoagulation therapy such as coumadin ® or warfarin.At this time there is no reason to suspect a malfunction exists.The reporting decision was based on limited information available that suggests the product was not performing within the variability of the assay.The investigation is underway.
|
| |
|
Additional Manufacturer Narrative
|
|
Apoc incident (b)(4).Apoc labeling will be evaluated during the investigation as pertaining to the event.
|
| |
|
Additional Manufacturer Narrative
|
|
Apoc incident: # (b)(4).The investigation was completed on 15-aug-2023.A review of the device history record (dhr) confirmed the cartridge lot passed release specifications.Retained cartridge testing of pt/inr cartridge lot t23054 did not meet the acceptance criteria in appendix 1 of q04.01.003 rev.Al (product complaint level 2 and level 3 investigation procedure) for points outside allowable error (ea).Returned cartridge testing meets the acceptance criteria in appendix 1 of q04.01.003 rev.Al (product complaint level 2 and level 3 investigation procedure).A deficiency has been identified for pt/inr cartridge lot t23054.Pt/inr cartridge lot t23054 has been added to quality record (qr) 899976 to address the root cause.
|
| |
|
Search Alerts/Recalls
|