| Model Number |
1080-212 |
| Medical Device Problem Code |
Mechanical Jam (2983)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/24/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that an impactor handle had a broken button mechanism.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned for investigation and it was confirmed the device was broken and the button would not move smoothly.A design change from a multi-component button design to a one-piece button has already been implemented.
|
| |
|
Search Alerts/Recalls
|