Model Number 6900PTFX33
Medical Device Problem Code Gradient Increase (1270)
Health Effect - Clinical Code Heart Failure/Congestive Heart Failure (4446)
Date of Event 05/20/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional manufacturer narrative: the investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Event or Problem Description
It was reported that a patient with a 33mm 6900ptfx pericardial valve in the tricuspid position underwent a valve-in-valve procedure after an approximate implant duration of four (4) years, nine (9) months due to tricuspid stenosis.The patient initially presented with heart failure and right ventricle dysfunction.The ttvr procedure was performed with a 29mm 9755rsl transcatheter valve.The patient was noted to be stable post procedure.
 
Additional Manufacturer Narrative
The device history record (dhr) review was completed and this device passed all manufacturing and sterilization inspections prior to release for distribution.Based on the information available, the device was intentionally used outside the way it was designed and intended.The performance, safety, efficacy, longevity, and durability of rings and bioprosthetic valves have not been determined when the product is used outside the scope of its intended use and/or its studied population.Stenosis, which develops progressively over time, can be due to a number of issues.Additionally, there can be a number of potential known and unknown patient-related contributing factors.Structural valve deterioration (svd) is the most common reason for bioprosthesis explant and encompasses multiple failure modes, including calcification, noncalcific degeneration, dehiscence, cusp thickening or fibrosis, or a combination of these.Such failure modes, occurring singularly or concomitantly, may contribute to stenosis and/or regurgitation.Alternatively, non-structural valve dysfunction (nsvd) may also play a role in the development of valvular stenosis.Stenosis is most commonly related to patient factors and is not usually an indication of a device malfunction.The most likely cause is patient factors, including end-stage renal disease (esrd) and implant position.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARPENTIER-EDWARDS PERIMOUNT THEON PERICARDIAL BIOPROSTHESIS
Common Device Name
REPLACEMENT HEART-VALVE
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer Contact
saurav singh
1 edwards way
irvine, CA 92614
9492506615
MDR Report Key17127376
Report Number2015691-2023-13687
Device Sequence Number7377032
Product Code DYE
UDI-Device Identifier00690103163629
UDI-Public(01)00690103163629(17)220128
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
P860057
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Expiration Date01/28/2022
Device Model Number6900PTFX33
Device Catalogue Number6900PTFX33
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/20/2023
Supplement Date Received by Manufacturer07/15/2023
Initial Report FDA Received Date06/14/2023
Supplement Report FDA Received Date07/24/2023
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention; Life Threatening;
Patient Age50 YR
Patient SexMale
-
-