| Model Number |
21-7381-24 |
| Medical Device Problem Code |
Inaccurate Delivery (2339)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that the customer had several complaints from patients about their infusion bags running out before the scheduled times and that the problem may be with the tubing.The cassette delivered 54 ml after programming for 50 ml.These did not meet a +/-6% threshold.All were tested with the same pump, which when sent in for preventative maintenance, was certified without issue.
|
| |
|
Additional Manufacturer Narrative
|
|
No product was returned.The reported complaint could not be confirmed.A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.If the product is returned this complaint will be reopened for further investigation.G1, email address: (b)(4).
|
| |
|