| Model Number |
5001102 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/15/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The sample has been received, but analysis of the sample has not yet been completed to date.Quest will file a supplemental report at the conclusion of the investigation.
|
| |
|
Event or Problem Description
|
|
It was reported to quest medical by (b)(6) of an alleged issue with product code 5001102.In the report, it was stated that blood was allegedly observed in the heat exchanger.It was confirmed that there were no patient complications.Lot number of the device was not reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was received by quest medical, inc.On 05/26/2023.Visual inspection confirmed the presence of blood on the top flange of the heat exchanger.It was also noted that the bellow within the hex had a deep indentation in its neck where the blood pooled.The root cause was inferred to be due to the deep neck during molding by the supplier, leading to inadequate glue application to prevent leakage.Quest will continue to monitor this complaint condition.Quest has concluded its investigation.
|
| |
|