MAUDE Adverse Event Report: QUEST MEDICAL, INC. MPS DELIVERY SET; CPBP HEAT EXCHANGER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

QUEST MEDICAL, INC. MPS DELIVERY SET; CPBP HEAT EXCHANGER Back to Search Results
Model Number 5001102
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/15/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The sample has been received, but analysis of the sample has not yet been completed to date.Quest will file a supplemental report at the conclusion of the investigation.
 
Event or Problem Description
It was reported to quest medical by (b)(6) of an alleged issue with product code 5001102.In the report, it was stated that blood was allegedly observed in the heat exchanger.It was confirmed that there were no patient complications.Lot number of the device was not reported.
 
Additional Manufacturer Narrative
The device was received by quest medical, inc.On 05/26/2023.Visual inspection confirmed the presence of blood on the top flange of the heat exchanger.It was also noted that the bellow within the hex had a deep indentation in its neck where the blood pooled.The root cause was inferred to be due to the deep neck during molding by the supplier, leading to inadequate glue application to prevent leakage.Quest will continue to monitor this complaint condition.Quest has concluded its investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MPS DELIVERY SET
Common Device Name
CPBP HEAT EXCHANGER
Manufacturer (Section D)
QUEST MEDICAL, INC.
1 allentown parkway
allen TX 75002
Manufacturer (Section G)
QUEST MEDICAL, INC.
1 allentown parkway
allen TX 75002
Manufacturer Contact
brian hedden
1 allentown parkway
allen, TX 75002
9723326311
MDR Report Key17128549
Report Number1649914-2023-00018
Device Sequence Number17754473
Product Code DTR
UDI-Device Identifier00634624501126
UDI-Public00634624501126
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K953838
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number5001102
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/26/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/15/2023
Supplement Date Received by Manufacturer05/15/2023
Initial Report FDA Received Date06/14/2023
Supplement Report FDA Received Date09/14/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-