• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH VA-LCP CLAVIC PL 2.7 LAT CS2 LE TAN; PLATE, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH VA-LCP CLAVIC PL 2.7 LAT CS2 LE TAN; PLATE, FIXATION, BONE Back to Search Results
Catalog Number 04.112.612S
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code Fall (1848)
Date of Event 05/29/2023
Type of Reportable Event Serious Injury
Event or Problem Description
Device report from synthes reports an event in japan as follows: it was reported that on (b)(6) 2023, the patient underwent orif surgery for distal clavicle fracture with clavicle plate and screws.The surgeon fixed the plate with six distal and four proximal screws.And it was reported that the screw of the distal clavicle plate was broken and was to be removed and replaced with a hook plate.The surgeon commented that it was not caused by the implants because the patient rode a bicycle despite his prohibition and fell, striking his shoulder.X-rays showed that the two most distal row screws were broken off and the plate was dislodged upward from the distal bone fragments.The proximal screws were unchanged.The clavicle plate will be removed on (b)(6) 2023, and revision surgery will be performed with another clavicle plate.The surgery was completed successfully without any surgical delay.Patient status or outcome is stable.No further information is available.This report is for one (1)va-lcp clavic pl 2.7 lat cs2 le tan.This is report 3 of 3 for complaint (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d9: complainant part is not expected to be returned for manufacturer review/investigation.Release to warehouse date: (b)(6) 2022.Expiration date: (b)(4) 2032.Part number: 04.112.612s, ti segmental shaft plate lateral/left/medium-sterile.Lot number: 667p951 (sterile).Note: plate was manufactured by elmira; lot number 637p545.Parts were packaged, sterilized and released by monument.Production order traveler met all inspection acceptance criteria.Packaging label log was reviewed and determined to be conforming.Packaging bom was reviewed and determined to be conforming with no deviations to normal packaging identified.Supplied was reviewed and determined to be conforming.This lot met all visual, sterility and packaging criteria at the time of release with no issues documented during the manufacture that would contribute to this complaint condition.Product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VA-LCP CLAVIC PL 2.7 LAT CS2 LE TAN
Common Device Name
PLATE, FIXATION, BONE
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 
3035526892
MDR Report Key17128958
Report Number8030965-2023-07466
Device Sequence Number7377494
Product Code HRS
UDI-Device Identifier07612334214294
UDI-Public(01)07612334214294
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K203414
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number04.112.612S
Device Lot Number667P951
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/29/2023
Initial Report FDA Received Date06/14/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/07/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNK - SCREWS: TRAUMA.; UNK - SCREWS: TRAUMA.; UNK - SCREWS: TRAUMA.; UNK - SCREWS: TRAUMA.; UNK - SCREWS: TRAUMA.; UNK - SCREWS: TRAUMA.; UNK - SCREWS: TRAUMA.; UNK - SCREWS: TRAUMA.; UNK - SCREWS: TRAUMA.; UNK - SCREWS: TRAUMA.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age60 YR
Patient SexMale
Patient Weight65 KG
-
-