| Model Number |
X SERIES |
| Medical Device Problem Code |
Unable to Obtain Readings (1516)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/23/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that while attempting to monitor a patient (age & gender unknown) the device displayed an "ecg monitoring failure" error message.Complainant indicated that the clinician obtained another device to continue treating the patient.Complainant indicate that there was no adverse effect to the patient due to the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation for evaluation.The customer's report was observed during review of the device data logs.However, the device was put through extensive testing including bench handling and ecg stress testing without duplicating the report.The multifunction cable (mfc) and defib receptacle were replaced as a precaution.The device was recertified and returned to the customer.The customer's mfc was returned for evaluation; no discrepancies were found during visual inspection.Analysis of reports of this type has not identified an increase in trend.
|
| |
|
Search Alerts/Recalls
|