| Model Number |
55-07764 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/27/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the plates were discovered to be broken and a revision surgery was performed.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has not yet been received at the manufacturer for testing.An evaluation will be conducted upon receipt of the device, and a follow-up report will be submitted after the quality investigation is complete.
|
| |
|
Event or Problem Description
|
|
It was reported that the plates were discovered to be broken and a revision surgery was performed.
|
| |
|
Additional Manufacturer Narrative
|
|
Update: h6 the reported event could be confirmed: the returned plate shows a fracture.The macroscopic appearances of the plate point to exceeding the material strength.That is confirmed by the deterioration of the anodization layer, the changes of the surface¿s structure and the secondary crack in the area of the breakages.It was further reported that bone healing was not achieved.Related to the event in question, a review of additional documentation (x-rays) was performed.These documents and information provided were forwarded to stryker¿s senior global medical director along with the complaint information.Upon review, he stated that to his opinion the plate was installed properly.In sum, the damages around the fractures indicate that the plate has been severely bent which decreased the material stability of the plate and led finally to the breakage.
|
| |
|