• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. BD INSYTE¿ AUTOGUARD¿ BC PRO SHIELDED IV CATHETER; INTRAVASCULAR CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. BD INSYTE¿ AUTOGUARD¿ BC PRO SHIELDED IV CATHETER; INTRAVASCULAR CATHETER Back to Search Results
Model Number 392523
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/16/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
B3.The date received by manufacturer has been used for this field.H.3.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that the bd insyte¿ autoguard¿ bc pro shielded iv catheter was backing out of the vein.The following information was provided by the initial reporter: verbatim: " during procedures the iv's go subcutaneous/infiltrate." additional information received on 22-may-2023.2.Could you please explain in detail the statement ¿iv go subcutaneous upon initiation of iv fluid in a vein that was patent¿ ? what does the iv go subcutaneous mean? - when initiating iv access with the catheters in question multiple nurses were using said catheters and iv's were patent when the iv was initiated in pre-op holding area and began at a slow keep vein open kvo rate once the patients were taken in the or or the procedure rooms and fluids were open wide open for infusion during procedures the iv's go subcutaneous/infiltrate - this was happening on multiple patients and iv were started by multiple nurses.
 
Additional Manufacturer Narrative
There was no sample or photo available to bd for evaluation.Therefore, bd was unable to perform a thorough investigation to verify the reported issue.Since, an investigation could not be performed bd was unable to determine a possible root cause.The manufacturing facility has been notified of this incident and the findings.A review of the device history record was performed and no quality issues were found during production.We appreciate you taking the time to bring this observation to our attention.We regret any inconveniences this incident may have caused you and your facility.Complaints received for this product and condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business regularly reviews the collected data for identification of emerging trends.
 
Event or Problem Description
No additional information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD INSYTE¿ AUTOGUARD¿ BC PRO SHIELDED IV CATHETER
Common Device Name
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
9450 south state street
sandy UT 84070
Manufacturer (Section G)
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
9450 south state street
sandy UT 84070
Manufacturer Contact
phillip emmert
5859 farinon drive
san antonio, TX 78249
8448235433
MDR Report Key17130877
Report Number1710034-2023-00661
Device Sequence Number20663428
Product Code FOZ
UDI-Device Identifier00382903925230
UDI-Public(01)00382903925230
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K201075
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 07/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2025
Device Model Number392523
Device Catalogue Number392523
Device Lot Number2308068
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/19/2023
Supplement Date Received by Manufacturer06/14/2023
Initial Report FDA Received Date06/14/2023
Supplement Report FDA Received Date07/23/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-