MAUDE Adverse Event Report: SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX BLUE LINA ULTRA SUCTIONAID TRACH TUBE; TUBE, TRACHEOSTOMY (W/WO CONNECTOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX BLUE LINA ULTRA SUCTIONAID TRACH TUBE; TUBE, TRACHEOSTOMY (W/WO CONNECTOR) Back to Search Results
Model Number 100/870/080
Medical Device Problem Code Material Puncture/Hole (1504)
Health Effect - Clinical Code Insufficient Information (4580)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported, that the doctor in charge of the procedure felt that the balloon came with holes.It's unknown, if there was patient involvement or any injury.
 
Additional Manufacturer Narrative
B3: date of event is unknown.No information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when required.
 
Additional Manufacturer Narrative
Other text: d10.Device available for evaluation; h3.Device evaluated by manufacturer and h6.Evaluation codes: updated.One decontaminated device was received.No visual observations were noted.A functional inflation test was performed in which it was not possible to inflate the cuff as it was leaking from a tear in the cuff.The root cause was due to manufacturing.A device history record (dhr) review reported no discrepancies or non-conformances during the manufacturing of the reported lot number.Corrective actions have been taken including initiation of quality alert, training of production, replacement of gauges and equipment with blunt versions, warning in manufacturing procedure and establishing better organization on affected workplace.
 
Additional Manufacturer Narrative
Device evaluation: no product sample was received; therefore, visual and functional testing could not be performed.The reported issue could not be confirmed.If the product is returned, the manufacturer will reopen this complaint for further investigation.A review of the device history records (dhr) shows there were no observations recorded during manufacture to suggest an issue of this nature would occur with this lot of product.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX BLUE LINA ULTRA SUCTIONAID TRACH TUBE
Common Device Name
TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
olomoucka 306
hranice 1, mesto 753 0 1
EZ  753 01
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key17131147
Report Number3012307300-2023-06254
Device Sequence Number12621053
Product Code BTO
UDI-Device Identifier15019315104862
UDI-Public15019315104862
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K030570
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number100/870/080
Device Catalogue Number100/870/080CZ
Device Lot Number4245254
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/30/2023
Supplement Date Received by Manufacturer06/23/2023
08/03/2023
Initial Report FDA Received Date06/14/2023
Supplement Report FDA Received Date06/26/2023
09/29/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/07/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-