| Model Number |
720054-02 |
| Medical Device Problem Code |
Material Discolored (1170)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/19/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported, that the patient underwent a procedure.To remove the ams spectra concealable penile prosthesis, because the device was bending.The device was replaced with a tactra malleable penile prothesis.There were no patient complications reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon receipt at our quality assurance laboratory, the returned device underwent a thorough analysis.Noth cylinders were visually inspected, and functionally tested.Visual examination revealed that both cylinders had holes in the distal of their bodies with internal component exposed, consistent with explant damage.In addition, it was noted that there were yellow specks on the body of cylinder 1.Both cylinders passed the functional inspection.Based on the information available and analysis results, the reported clinical observation of device bending could not be confirmed; therefore, a conclusion code of cause not established was assigned to this investigation.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent a procedure to remove the ams spectra concealable penile prosthesis because the device was bending.The device was replaced with a tactra malleable penile prothesis.There were no patient complications reported.
|
| |
|