|
H6: investigation summary photos received by our quality team for investigation.Through visual inspection of the picture, the protector is assembled in the vial, but the rubber stopper has been disassembled from the vial and has fallen into the medicine vial.Through visual evaluation of the samples, no defects or issues observed.Functional testing was performed, connecting the protector sample to both a vial, injector, and syringe per the instructions for use provided with the product.In all cases the expansion chamber expanded properly, liquid was able to be drawn from the vial, and the product functioned as intended.A device history review was performed for lot 2111401, no deviations or non-conformances were identified during the manufacturing process that could have contributed to the reported issue.Three retained samples from the lot were evaluated, no defects or issues observed.Functional testing was performed, connecting the protector sample to both a vial, injector, and syringe per the instructions for use provided with the product.In all cases the expansion chamber expanded properly, liquid was able to be drawn from the vial, and the product functioned as intended.Product undergoes a series of testing during manufacturing to ensure the quality and functionality of the device.These tests include visual inspections along with leakage testing and flow rate, all records were reviewed for the reported lot and results were found to be acceptable.Based on the quality team's investigation, we are not able to identify a root cause related to our manufacturing process at this time.
|