| Model Number |
S602 |
| Medical Device Problem Code |
Battery Problem (2885)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/09/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation determined that battery status indicators were different than expected based on information provided to boston scientific.Factors influencing the estimated longevity remaining calculation include pacing rate, amplitude, pulse-width and lead impedance.Any (even slight) changes in these factors will impact the battery consumption calculation and therefore the remaining longevity estimate.Please refer to the description for more information regarding the specific circumstances of this event.Please note this pacemaker's battery was designed to estimate remaining longevity (and trigger corresponding device replacement indicators) based on the pacing capacitor charge time, which progressively increases as the battery depletes.These pacemakers were not designed with an ability to calculate remaining longevity for every possible current drain, which may cause replacement indicators to trigger earlier than what was previously estimated; however, it does not negatively impact the use, operation, safety, or performance of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that this implantable pacemaker longevity three months ago was one and a half year remaining and recently longevity was noted at more than five years with output is auto 5 volts.Boston scientific technical services (ts) reviewed and discussed that device diagnostics appears normal.This is possibly device reset that clears battery, device logbook and daily measurements.Moreover, the auto output was likely associated with the reset.Ts then suggested to bring patient in to test threshold and program appropriately.Furthermore, ts suggested to replace device if diagnostic have not returned in two to four weeks.The device remains in service and has been implanted for almost 15 years.No adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation determined that battery status indicators were different than expected based on information provided to boston scientific.Factors influencing the estimated longevity remaining calculation include pacing rate, amplitude, pulse-width and lead impedance.Any (even slight) changes in these factors will impact the battery consumption calculation and therefore the remaining longevity estimate.Please refer to the description for more information regarding the specific circumstances of this event.Please note this pacemaker's battery was designed to estimate remaining longevity (and trigger corresponding device replacement indicators) based on the pacing capacitor charge time, which progressively increases as the battery depletes.These pacemakers were not designed with an ability to calculate remaining longevity for every possible current drain, which may cause replacement indicators to trigger earlier than what was previously estimated; however, it does not negatively impact the use, operation, safety, or performance of the device.This product has not been returned for analysis.This report will be updated should additional information become available or if analysis is completed.
|
| |
|
Event or Problem Description
|
|
It was reported that this implantable pacemaker longevity three months ago was one and a half year remaining and recently longevity was noted at more than five years with output is auto 5 volts.Boston scientific technical services (ts) reviewed and discussed that device diagnostics appears normal.This is possibly device reset that clears battery, device logbook and daily measurements.Moreover, the auto output was likely associated with the reset.Ts then suggested to bring patient in to test threshold and program appropriately.Furthermore, ts suggested to replace device if diagnostic have not returned in two to four weeks.The device remains in service and has been implanted for almost 15 years.No adverse patient effects were reported.Additional information indicated that the device was explanted and replaced.No additional adverse patient effects were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that this implantable pacemaker longevity three months ago was one and a half year remaining and recently longevity was noted at more than five years with output is auto 5 volts.Boston scientific technical services (ts) reviewed and discussed that device diagnostics appears normal.This is possibly device reset that clears battery, device logbook and daily measurements.Moreover, the auto output was likely associated with the reset.Ts then suggested to bring patient in to test threshold and program appropriately.Furthermore, ts suggested to replace device if diagnostic have not returned in two to four weeks.The device remains in service and has been implanted for almost 15 years.No adverse patient effects were reported.Additional information indicated that the device was explanted and replaced.No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon receipt at our post market quality assurance laboratory, detailed mechanical and electrical testing was performed on the device.The device battery status was not as expected.The device functioned normally throughout testing.Laboratory analysis determined that this device experienced normal battery depletion; however, the estimated longevity remaining value appeared to decrease more quickly than expected between routine follow-ups.Factors influencing the estimated longevity remaining calculation include pacing rate, amplitude, pulse-width and lead impedance.Any (even slight) changes in these factors will impact the battery consumption calculation and therefore the remaining longevity estimate.Please note that, despite the drop in estimated longevity remaining, the actual battery condition did not change significantly between follow-ups.In summary, it was determined that this device experienced normal battery depletion, but declared eol earlier than previously estimated.Of note, the estimated remaining battery longevity of this pacemaker was designed to be calculated (and trigger corresponding device replacement indicators) based on the pacing capacitor charge time, which progressively increases as the battery depletes.These pacemakers were not designed with an ability to calculate remaining longevity for every possible current drain, which may cause replacement indicators to trigger earlier than what was previously estimated; however, it does not negatively impact the use, operation, safety, or performance of the device.This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation determined that battery status indicators were different than expected based on information provided to boston scientific.Factors influencing the estimated longevity remaining calculation include pacing rate, amplitude, pulse-width and lead impedance.Any (even slight) changes in these factors will impact the battery consumption calculation and therefore the remaining longevity estimate.Please refer to the description for more information regarding the specific circumstances of this event.Please note this pacemaker's battery was designed to estimate remaining longevity (and trigger corresponding device replacement indicators) based on the pacing capacitor charge time, which progressively increases as the battery depletes.These pacemakers were not designed with an ability to calculate remaining longevity for every possible current drain, which may cause replacement indicators to trigger earlier than what was previously estimated; however, it does not negatively impact the use, operation, safety, or performance of the device.This product has not been returned for analysis.This report will be updated should additional information become available or if analysis is completed.
|
| |
|