• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON GRIESHABER AG FINESSE SHARKSKIN ILM FORCEPS; FORCEPS, OPHTHALMIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON GRIESHABER AG FINESSE SHARKSKIN ILM FORCEPS; FORCEPS, OPHTHALMIC Back to Search Results
Catalog Number 705.88P
Medical Device Problem Code Difficult to Open or Close (2921)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/06/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A nurse reported that during a vitrectomy surgery an ophthalmic forceps did not work.Surgery was completed by an alternate forcep.No patient harm was reported.
 
Additional Manufacturer Narrative
The sample is received with inner blister, outer blister and cover foil showing the lot information.The sample shows no macroscopic signs of damage.The sample shows signs of surgery residues and is returned in a opened status.The sample was visually inspected with the aid of a photomicroscope with various magnifications.The sample was functionally tested.It was noticed that the forceps get stuck in a closed status.The forceps only can get open again by pushing the shaft manually down.As the blister was opened by the customer, he handled the product.The point in time when the malfunction occurred, can no longer be determent.It cannot be determined how or where the damage to the sample occurred; therefore a root cause for the customers reported event could not be determined.However, the damage was consistent with damage that can occur due to improper handling.The exact root cause for the customers reported event is unknown, therefore, specific action cannot be taken the manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FINESSE SHARKSKIN ILM FORCEPS
Common Device Name
FORCEPS, OPHTHALMIC
Manufacturer (Section D)
ALCON GRIESHABER AG
winkelriedstrasse 52
schaffhausen 8203
SZ  8203
Manufacturer (Section G)
ALCON GRIESHABER AG
winkelriedstrasse 52
schaffhausen 8203
SZ   8203
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key17133560
Report Number3003398873-2023-00068
Device Sequence Number6448613
Product Code HNR
UDI-Device Identifier07612717071544
UDI-Public07612717071544
Combination Product (Y/N)N
Initial Reporter CountryFI
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Nurse
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 09/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2025
Device Catalogue Number705.88P
Device Lot Number14AVMX
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/04/2023
Supplement Date Received by Manufacturer08/23/2023
Initial Report FDA Received Date06/15/2023
Supplement Report FDA Received Date09/14/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/22/2022
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-