| Catalog Number |
824M6545 |
| Medical Device Problem Code |
Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
05/22/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.Reference reports 3012447612-2023-00197 through 3012447612-2023-00198.
|
| |
|
Event or Problem Description
|
|
It was reported that 12 days after surgery, the patient heard a noise from the implants and an x-ray found that the closure top and rod had migrated out of the pedicle screw on the left side of l5.A revision surgery was performed to remove the pedicle screw, rod, and closure top and replace them with another pedicle screw, rod, and closure top.There were no injuries that occurred during the revision surgery.This is report three of three for this event.
|
| |
|
Event or Problem Description
|
|
It was reported that 12 days after surgery, the patient heard a noise from the implants and an x-ray found that the closure top and rod had migrated out of the pedicle screw on the left side of l5.A revision surgery was performed to remove the pedicle screw, rod, and closure top and replace them with another pedicle screw, rod, and closure top.There were no injuries that occurred during the revision surgery.This is report three of three for this event.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: additional method codes: 4109 and 4110.Device evaluation: the returned device matches the information in the complaint file and was examined.Visual inspection revealed no clear damage to the device and all the threads were intact.X-rays were provided showing that the closure top and rod backed out of the pedicle screw.Root cause: a definitive root cause cannot be determined with the information provided.This event could possibly be attributed to unknown patient or surgical factors.Dhr review: the dhr was reviewed.There are no indications of manufacturing issues which would have contributed to this event and the device was likely conforming when it left highridge medical¿s control.Device usage: this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
|
| |
|