• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON CLAREON ASPHERIC HYDROPHOBIC ACRYLIC IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON CLAREON ASPHERIC HYDROPHOBIC ACRYLIC IOL; INTRAOCULAR LENS Back to Search Results
Model Number SY60WF
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Conjunctivitis (1784); Endophthalmitis (1835); Vitreous Floaters (1866); Hypopyon (1913); Blurred Vision (2137); Fibrosis (3167); Eye Pain (4467); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 05/08/2023
Type of Reportable Event Serious Injury
Event or Problem Description
A nurse reported that after postoperative day 14 of cataract surgery with ophthalmic handpiece, intraocular lens, phacoemulsification tip, viscoelastic device and consumable, a 68 years old female patient experienced with endophthalmitis in left eye with symptoms of eye pain, blurry vision, floater, conjunctival inflammation was <1+, aqueous cell was 3+, aqueous fibrin was present with hypopyon.Patient was treated with intravitreal tap and antibiotics, corticosteroids medications.Culture results revealed no bacterial growth.Current patient condition was unknown.This is the report of patient 1 of 4.
 
Additional Manufacturer Narrative
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
The product was not returned.Product history records were reviewed and documentation indicated the product met release criteria.The product investigation could not identify a root cause for the reported complaint.The lens remains implanted.All iols are terminally sterilized with 100% ethylene oxide sterilization.Company uses an overkill sterilization approach, which greatly exceeds the minimum requirements for sterility.Critical parameters for sterilization cycles are recorded and retained for every sterilization load to demonstrate that the acceptance criteria for the sterilization process are met.Information was provided by the reporter that it was unknown what may have caused the event.The patient's vision was indicated to be improving.No further information has been provided.The investigation has been completed based on current information.Based on our current tracking, there are no adverse trends for this reported complaint.No further action warranted at this time.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAREON ASPHERIC HYDROPHOBIC ACRYLIC IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key17135126
Report Number1119421-2023-01120
Device Sequence Number14945450
Product Code HQL
UDI-Device Identifier00380652251211
UDI-Public00380652251211
Combination Product (Y/N)N
Initial Reporter StateND
Initial Reporter CountryUS
PMA/510(K) Number
P190018
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/01/2025
Device Model NumberSY60WF
Device Catalogue NumberSY60WF.065
Device Lot Number15301431
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/17/2023
Supplement Date Received by Manufacturer08/21/2023
Initial Report FDA Received Date06/15/2023
Supplement Report FDA Received Date09/14/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/08/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CEFTADAZIME; DEXAMETHASONE; VANCOMYCIN
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient Age68 YR
Patient SexFemale
-
-