Model Number DFR00V
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Halo (2227); Insufficient Information (4580); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 06/28/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Section a4 and a5: unknown; requested but not provided.Section h3-other (81): the device was not returned for evaluation; therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.Section h6: health effect - clinical code: code 4581 is to capture device decentered or dislocated or tilted subluxated or wrong position, and code 4580 is to capture "balance issues".Attempts were made to obtain the missing information; however, no additional information has been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that the intraocular lens (iol) was explanted from the left eye due to dislocation, halos, and balance issues.The patient did not have a debilitating condition due to these issues.The patient does not have any pre-existing medical history issues.A non-johnson and johnson light adjustable lens (lal) was implanted.No additional surgical interventions were required.There was no patient injury.The patient is ¿ok¿ post-operatively.No further information was provided.
 
Additional Manufacturer Narrative
The product was received for evaluation.Additional information: section d9: device available for evaluation? yes.Section d9: date returned to manufacturer: jul 26, 2023.Section h3: evaluated by manufacturer: yes.Device evaluation: two lenses were received for evaluation with no identifying information; therefore, both lenses were evaluated and the results are provided in this report.Sample 1: visual inspection under magnification revealed that the complaint lens was received cut in half (one half missing).The lens was cleaned and, no issues that could cause or contribute to the complaint issue could be identified.Based on the return condition of the lens, no further product evaluation could be performed.The complaint issue of ¿instability of device after implantation¿ was not confirmed.The other observed issue found during the product evaluation could not be confirmed to be related to a manufacturing or design issue.Sample 2: visual inspection under magnification revealed that the complaint lens was received cut in half.The lens was cleaned and, no issues that could cause or contribute to the complaint issue could be identified.Based on the return.The complaint issue of ¿instability of device after implantation¿ was not confirmed.The other observed issue found during the product evaluation could not be confirmed to be related to a manufacturing or design issue.Conclusion: based on the complaint investigation results, the product was released within specifications.The complaint issue reported could not be confirmed and no product deficiency or product malfunction could be identified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
LENS, MULTIFOCAL INTRAOCULAR
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key17135960
Report Number3012236936-2023-01497
Device Sequence Number20338760
Product Code MFK
UDI-Device Identifier05050474750548
UDI-Public(01)05050474750548(17)240403
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/03/2024
Device Model NumberDFR00V
Device Catalogue NumberDFR00VU135
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/23/2023
Supplement Date Received by Manufacturer08/29/2023
Initial Report FDA Received Date06/15/2023
Supplement Report FDA Received Date09/11/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/03/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
Patient RaceWhite
-
-