| Model Number |
T4 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/03/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The analyzer serial number is (b)(6).An interfering substance is suspected in the patient sample.The patient sample was requested for investigation.The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
There was an allegation of questionable elecsys t4 assay results for 1 patient sample on a cobas e 801 analytical unit compared to a beckman analyzer.Refer to the attachment to the medwatch for all patient data.Clarification on if the results from (b)(6) 2023 were from a new sample was requested but not provided.The results were reported outside of the laboratory.
|
| |
|
Additional Manufacturer Narrative
|
|
The beckman analyzer results were produced on (b)(6) 2023.The results from (b)(6) 2023 were obtained from the same sample.The qc recovery data provided was acceptable.A general product problem could be excluded.The investigation found that the patient sample contains an interfering factor to the streptavidin component of the reagent.Per product labeling, "in rare cases, interference due to extremely high titers of antibodies to analyte-specific antibodies, streptavidin or ruthenium can occur.These effects are minimized by suitable test design.".
|
| |
|