• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEVILBISS HEALTHCARE LLC DEVILBISS; SUCTION UNIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEVILBISS HEALTHCARE LLC DEVILBISS; SUCTION UNIT Back to Search Results
Model Number 7314P-D
Medical Device Problem Code Suction Problem (2170)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
Devilbiss healthcare was notified of an incident involving a suction unit by a provider, who stated the "unit has no suction." there was no report or evidence of illness, injury or medical treatment associated with the complaint.The device was returned to devilbiss for evaluation, where it was determined the unit was unable to produce vacuum pressure to specification.The root cause was the umbrella-style check valve that dislodged from the compressor cylinder and a loose manifold.Devilbiss addressed check valve complaints via capa.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEVILBISS
Common Device Name
SUCTION UNIT
Manufacturer (Section D)
DEVILBISS HEALTHCARE LLC
100 devilbiss drive
somerset PA 15501
Manufacturer (Section G)
DEVILBISS HEALTHCARE LLC
100 devilbiss drive
somerset PA 15501
Manufacturer Contact
roberto munoz
100 devilbiss drive
somerset, PA 15501
5169984600
MDR Report Key17136438
Report Number2515872-2023-00065
Device Sequence Number5696007
Product Code JCX
UDI-Device Identifier00885304000327
UDI-Public885304000327
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K982304
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other
Remedial Action Repair,Inspection
Type of Report Initial
Report Date (Section B) 06/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number7314P-D
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/15/2023
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/01/2023
Initial Report FDA Received Date06/15/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/13/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-