• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Model Number 3542257
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Deformity/ Disfigurement (2360)
Date of Event 04/18/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
At the time of this report, mentor has received no information regarding explantation or an expected explantation date.It is unknown at this time if the device will be made available for return.As a result, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: no information regarding explantation or an expected explantation date has been received.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
 
Event or Problem Description
It was reported that a 70-year-old female patient underwent a primary breast augmentation with a 225cc mentor memorygel breast implant and experienced a rupture on the right side post-operatively.At the time of this report, mentor has received no information regarding explantation or an expected explantation date.
 
Additional Manufacturer Narrative
On august 16, 2023, mentor received a new date of explant.Relevant fields have been updated.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On september 18, 2023, the mentor failure analysis lab received the device for evaluation.Device evaluation summary: the product was returned to mentor for evaluation.Mentor conducted visual inspection of the returned device.Visual analysis of the returned sample revealed that the gel siltex mod-rnd 225cc breast implant had a tear on the posterior view, measuring approximately 1.3 cm.In addition, an area of siltex cracking was found at the edge of the tear.The evaluation determined that the possible cause of the rupture is consistent with normal wear.Siltex cracking is defined as a material separation occurring in the siltex layer and can eventually progress through the shell of siltex devices.Siltex cracking is ultimately a result of high stresses.As part of mentor¿s quality process all devices are manufactured, inspected, and released to approved specifications.Capa activity was not required.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post market surveillance.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR MEMORYGEL BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
kate karberg
3041 skyway circle north
irving, TX 75038
3035526892
MDR Report Key17139643
Report Number1645337-2023-07164
Device Sequence Number5696110
Product Code FTR
UDI-Device Identifier00081317000648
UDI-Public00081317000648
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P030053
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2001
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2006
Device Model Number3542257
Device Catalogue Number3542257
Device Lot Number211050
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/26/2023
Supplement Date Received by Manufacturer08/16/2023
09/18/2023
Initial Report FDA Received Date06/15/2023
Supplement Report FDA Received Date09/05/2023
09/21/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/01/2000
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient Age70 YR
Patient SexFemale
-
-