| Model Number |
11-4090 |
| Medical Device Problem Code |
Separation Failure (2547)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
03/31/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The removed rod and locking screws were returned for evaluation.Evaluation is in progress.Review of production records identified no issues related to production or inspection that may be associated to the reported event.
|
| |
|
Event or Problem Description
|
|
At an unspecified time post-operatively, a surgical procedure was performed to address a loose rod in a pedicle screw construct.One rod and four locking screws were removed and replaced at that time.
|
| |
|
Event or Problem Description
|
|
Four days post-operatively prior to patient discharge, one loose locking screw was identified radiographically.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: mfg narrative update: the removed rod and locking screws were returned and evaluated.Visual inspection identified marking and deformation consistent with the reported event.Based on evaluation findings and the information reported, the exact root cause could not be determined.
|
| |
|