| Model Number |
BB811 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/15/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Medtronic received information that after the use of a fusion oxygenator, the customer observed a leak from the temperature monitoring adapter (tma).This was observed after the case during tear down.There was no patient impact associated with this event.There was no patient blood loss due to this issue.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information b5: medtronic received additional information that the customer allegedly found this leak during tear down after the case, so the customer has assumed that there was some manipulation done to the tma at that time.Device evaluation summary: visual inspection showed that the temperature monitoring adapter (tma) was loose, and it could be easily removed.Additional visual inspection showed evidence of adhesive.The reason for return was confirmed.Conclusion: the complaint was confirmed for the fusion oxygenator's loose temperature monitoring adapter (tma).The issue was verified via analysis of the returned device.Review of this unit¿s device history record found no abnormalities or non-conforming material reports (ncmrs) initiated during manufacturing that would cause or contribute to the reported event.This unit passed all required testing and inspections during manufacturing.Customer communication of this issue has been initiated via fa1302, which provides device use recommendations for the affected product.Root cause investigation concluded the presence of ethylene bis-stearamide (ebs) wax on the new tma component affected the material properties that caused the torque specification to not be met.There were no patient/clinical safety issues reported.Trends for issues with this product are reviewed at quarterly quality meetings.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|