| Model Number |
MAJ-2117 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/02/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has not been returned for evaluation.If additional information becomes available prior to the conclusion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
The customer reported that his olympus connecting tubes were breaking during reprocessing.According to the initial reporter, the connector¿s right and left legs were falling off.There was no patient involvement during this event.For reference to related events, please see reports with the following patient identifiers: (b)(6), (b)(6), (b)(6), (b)(6), (b0(6), (b)(6), (b)(6), (b)(6).
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The device history record was unable to be reviewed for this device since the serial number was not provided.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.Based on the results of the investigation, and because the device was not received for evaluation, the definitive root cause of the broken connecting tube could not be determined.It is possible that foreign material was caught between the connection, or that external stress was applied to the tube which broke the lock lever and caused the user to apply force toward the incorrect direction.The instruction manual identifies the following verbiage, which may have prevented the phenomenon: ¿visually inspect the connecting tube to ensure that there are no cracks, breaks, rips, scratches, attached debris, or other irregularities.Move the lock levers of the reprocessor side connector to make sure that they function properly and are not broken.¿ olympus will continue to monitor field performance for this device.
|
| |
|
Search Alerts/Recalls
|
|
|