• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER LEIBINGER FREIBURG L-PLATE, 100 DEGREE, 8MM BAR, RIGHT, GSP, 6 HOLE, MIDFACE; IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER LEIBINGER FREIBURG L-PLATE, 100 DEGREE, 8MM BAR, RIGHT, GSP, 6 HOLE, MIDFACE; IMPLANT Back to Search Results
Model Number 55-07765
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/10/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the plate was broken and a revision surgery was performed.
 
Additional Manufacturer Narrative
The device has not yet been received at the manufacturer for testing.An evaluation will be conducted upon receipt of the device, and a follow-up report will be submitted after the quality investigation is complete.
 
Additional Manufacturer Narrative
Corrected data: the devices were returned to the manufacturing site for inspection.Upon inspection, the reported device was not found broken but cut.As a result, the plate is considered concomitant.It did not cause or contribute to the reported event (broken plate).
 
Event or Problem Description
It was reported that the plate was broken and a revision surgery was performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
L-PLATE, 100 DEGREE, 8MM BAR, RIGHT, GSP, 6 HOLE, MIDFACE
Common Device Name
IMPLANT
Manufacturer (Section D)
STRYKER LEIBINGER FREIBURG
boetzingerstr. 41
freiburg D-791 11
GM  D-79111
Manufacturer (Section G)
STRYKER LEIBINGER FREIBURG
boetzingerstr. 41
freiburg D-791 11
GM   D-79111
Manufacturer Contact
rylee koole
boetzingerstr. 41
freiburg D-791-11
GM   D-79111
7614888208
MDR Report Key17144220
Report Number0008010177-2023-00038
Device Sequence Number5589724
Product Code JEY
UDI-Device Identifier04546540366146
UDI-Public04546540366146
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Model Number55-07765
Device Catalogue Number55-07765
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/19/2023
Supplement Date Received by Manufacturer05/19/2023
Initial Report FDA Received Date06/16/2023
Supplement Report FDA Received Date08/07/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-