Model Number J20007D
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Corneal Clouding/Hazing (1878)
Date of Event 05/19/2023
Type of Reportable Event Serious Injury
Event or Problem Description
Customer reported that a patient presented with mild central corneal haze after lasik treatment.The haze resolved with the normal post operative steroid use.Physician has had multiple lasik cases without corneal haze.Physician plans to review the site¿s instrument preparation/sterilization process.Per follow-up, physician cannot remember the patients.There were only 2, and i have not seen anyone else since then.Patient was given prednisolone acetate - the normal drop for post op, but physician increased the frequency to 6x/day instead of 4x/day to hopefully help with the haze.This report is for patient 2 of 2.
 
Additional Manufacturer Narrative
Section h3-other (81): the intralase femtosecond system was not evaluated; therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.All pertinent information available to johnson & johnson surgical vision, inc has been submitted.H3 other text : see h10.
 
Additional Manufacturer Narrative
H10, corrected data: h4 ¿ manufacturing date ¿ the manufacturing site reported that the manufacturing date for the device is 05/20/2020.H10, additional information: h6 type of investigation codes 3331, 4110 and 4109.H6 investigation findings ¿ code 213.Device evaluation: a review of the records related to this equipment that included labeling, manual, trending, and risk documentation reviews for this equipment were performed.Equipment labeling provides possible complications that can be caused by the surgical/treatment procedure being performed.The trend review shows that there is not a recognizable adverse trend.The risks and mitigations associated with the complaint issue are identified in existing risk documents and no new risks were identified as part of this investigation.There was no product deficiency identified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IFS ADVANCED FEMTOSEC LASER
Common Device Name
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
AMO MANUFACTURING USA, LLC
510 cottonwood drive
milpitas CA 95035
Manufacturer Contact
somyata nagpal
1700 e st andrew place
santa ana, CA 92705
7142478552
MDR Report Key17146385
Report Number3012236936-2023-01523
Device Sequence Number5771958
Product Code GEX
UDI-Device Identifier05050474573468
UDI-Public(01)05050474573468
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K141852
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberJ20007D
Device Catalogue NumberJ20007D
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/19/2023
Supplement Date Received by Manufacturer06/23/2023
Initial Report FDA Received Date06/16/2023
Supplement Report FDA Received Date06/27/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/20/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-