Model Number RA002-4545SL
Medical Device Problem Codes Corroded (1131); Mechanical Jam (2983)
Health Effect - Clinical Code Failure of Implant (1924)
Type of Reportable Event Malfunction
Event or Problem Description
An investigation report from an implant retrieval center was received and reported that the rod would not extend and force testing was unable to be completed.Additionally, it was noted during surface damage grading that there was corrosion on the rod.X-ray of the rod showed metallic debris around the pin area.It was reported that the rod was removed as part of a planned removal procedure after desired lengthening had been achieved; there was no adverse patient or user impact associated with this report.No additional information is available.
 
Additional Manufacturer Narrative
The device has not been returned for evaluation.The evaluation was performed by the london implant retrieval center (lirc).The root cause is unable to be determined at this time.A review of the device history records (dhr) confirmed the device met all quality inspections and specifications prior to release.If any additional information is provided, a supplemental report will be submitted.Report 1 of 2.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGEC GENERATION 1.5 SPINAL BRACING AND DISTRACTION SYSTEM
Common Device Name
GROWING ROD SYSTEM - MAGNETIC ACTUATION
Manufacturer (Section D)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr. ste 100
aliso viejo CA 92656
Manufacturer (Section G)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr. ste 100
aliso viejo CA 92656
Manufacturer Contact
geoff gannon
101 enterprise dr. ste 100
aliso viejo, CA 92656
8583448112
MDR Report Key17148841
Report Number3006179046-2023-00289
Device Sequence Number5422010
Product Code PGN
UDI-Device Identifier00856719002183
UDI-Public856719002183
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K160352
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 06/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberRA002-4545SL
Device Lot NumberA170118-07
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/05/2023
Initial Report FDA Received Date06/16/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/18/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age7 YR
Patient SexFemale
-
-